- 27 -
7.3
Log interpretation
The format of the sterilization record for printer output and internal memory storage is as follows.
- - - - - - - - - - - - - - - -
Quantus St
erilizer
- - - - - - - - - - - - - - - -
Program : Quick-program-S
134°C
unwrapped
Data : 2017.07.12
Start Time : 09:11:29
Cycle : 1
SN : 65252120517050513
Preheat Temper : 95.0°C
Conductivity
:
11.0us/cm
- - - - - - - - - - - - - - - -
Program step press temper time
bar°C min
Start : -0.0469.6 0:00:00
1.Vacuum one
Sub press : -0.8772.00:02:26
Steam press: 0.50109.90:05:32
2.Vacuum two
Sub press : -0.5289.20:06:32
Steam press: 2.14134.30:11:51
- - - - - - - - - - - - - - - -
Keep : 2.14134.30:11:51
Keep End : 2.17135.0 0:16:51
Realse : 0.10 105.6 0:18:27
Dry : 0.10105.60:18:27
Dry End : -0.9190.3 0:22:27
Ventile : -0.1492.3 0:22:39
Finish : -0.1492.3 0:22:39
- - - - - - - - - - - - - - - -
PROGRAM PROPERLY EXECUTED!
Temperature : 134.7+0.2/-0.4°C
Pressure : 2.14+0.02/-0.03bar
Sterilization : 0
:
22:39
End Time : 09:34:08
Hardware Version: Qx3_V0.2
Software Version: 00.02.16.19
- - - - - - - - - - - - - - - -
Product Number
Program
Date
Start Time
Running times
Serial number
Preheating temperature
Distilled water conductivity
Vacuum twice, recording pressure,
Temperature time information
Sterilization holding stage
Sterilization holding completion
Depressurization
Drying start
Drying completion
Inlet
Sterilization completion
Final report, recording sterilization
temperature,pressure, program running time
and completion time
Hardware and software version