- 14 -
Table 4‐1 Sterilization program and parameter settings
Name of
Sterilization
T
(
℃
)
Holding
time
(Min)
Drying
time
(min)
Sterilization load
121
program
121
>
15
1‐99
a small amount of fabric
,
Heat‐resistant instrument.
134
program
134
>
3
1‐99
Mixed load
,
long and hollow
instruments
134+
program
134
>
3
1‐99
Mixed load
,
long and hollow
instruments
Quick S
134
>
3
1‐99
Solid instruments and hollow
instruments.
Quick B
134
>
3
1‐99
Wrapped & unwrapped solid
instruments
Fabric
121
>
15
1‐99
Fabric
Select the sterilization program as follows
:
Fig 4.1
Select sterilization program
4.3
Auxiliary option
4.3.1
Warm‐up setting
The automatic warm‐up function allows the sterilizer to reach the set temperature before the program is
started to significantly reduce the time required for each sterilization. See section 6.4 for details.
1 – Sterilization program
2 – Test program
2‐
3 ‐
System settings
4 – File record
Press
to select
134 program
121 program
Quick B
Quick S
Press
to enter