- 10 -
This product has water quality monitoring device,it will not work
properly when unqualified water and will give warning that water
quality cannot reach the standard.When you see this message,
please replace high quality distilled water in time.
If you need to use the water purifier, please consult the
manufacturer of this device first.
It is best to tilt and secure all the water pipes connected to the back
of the device are not bent or knotted.
2.4.3
Wastewater connection
For small steam sterilizers without internal water tank, take out the drain pipe from the accessory bag
and insert the waste water connection to the back of the device (Figure 1.4).
It is best to tilt and secure all the water pipes connected to the
back of the device are not bent or knotted.
2.5
Installation service list
After installation
please carefully fill out the installation service
form in the manual which is useful information for maintenance.
2.6
Handling instructions
This device has to be stored and transported at humidity less than 85% and temperature ‐20
o
C ~ 55
o
C.
Before transportation, please carry out “drainage program” first and take off all pipes(including distilled
water pipes and waste water pipes),if printer is connected,remove the printer cable.
If the tray rack and tray are placed in the chamber during transportation. Before closing the door,place a foam
pad between the door and the tray holder to prevent the tray frame from being damaged by shaking during
transportation
For long‐distance transportation, the product must be returned to the package in the order in which it was
unpacked, and the package must be sealed.
Please carry drainage program before handling this device