
- 11 -
3
First time operation
3.1
Startup
Turn on the power switch, the power switch position is in the lower right corner of the device inside view
(Figure 1.4).
Make sure to remove the latex plug from the air filter mounting
hole at the back of the device and install the air filter from the
accessories included.
3.2
After Startup
After startup at the first time, please set up language and carry out pressure test.
After the pressure verification
is successful, the initial setting is completed.
The date and time can also be set as required. See Chapter 6 system
settings in this manual.
Steps of pressure verification is as follows:
3.3
Open/close the door
Pressure verification
Verification is checking …
P: 0.00 Bar
Pressure verification
Pressure verification
successful
Door open to complete
pressure verification
press
button
and open up the door
Wait
for
calibration
and
make sure the
door is open.
press
to select
1–Sterilization program
2 – Test program
2‐
3 – Sytem setting
4 – File record