 Small steam sterilizer for sterilizing medical and laboratory equipments, The product design conforms 

to the YY/T 0646‐2015 standard and is a Class B sterilizer.Process evaluation system according to EN13060.

 

 

 

Do  not  sterilize  any  liquid

Violating  the  use  of  this  product  to 

sterilize liquids will damage the product and the loads.

 

 

 

 

   : 

Symbols 

Indication 

Description 

 

Warning

 

Identifies  danger  points  or  conditions  and 

the  correct  procedures  that  must  be 

followed to avoid injury. 

 

Hot Surfaces 

Hot items, take care to avoid burns. 

 

Caution 

Identifies  actions  that  could  damage  the 

device/component. 

 

Precautions

 

Draw user's attention to important matters.

 

 
 
 

     Symbols on products: 

 

Symbols 

Description 

 

take care to avoid burns 

 

Manufacturer of medical equipment

 

 

Date of manufacture of this device

 

 

Serial number 

Symbols description 

Application 

Summary of Contents for Q50B

Page 1: ...Small Steam Sterilizer User Manual Internal Water Tank Q50B Q70B 2460 H Tmed Wuxi Co Ltd V01 version...

Page 2: ...product and the loads Symbols Indication Description Warning Identifies danger points or conditions and the correct procedures that must be followed to avoid injury Hot Surfaces Hot items take care t...

Page 3: ...quid sterilization Do not sterilize any liquid Professionals The operator who use sterilizers and fabric must be got a trained professional Power cord and power plug Do not replace the power cable or...

Page 4: ...es abnormally at certain stages of the program high temperature steam will emerge from the chamber after opening the door The loaded item may not be sterilized when the sterilization is abnormally ter...

Page 5: ...f sterilization Standard 134 Standard 121 Quick B Quick S 134 Fabric Class of sterilization B B B S B B Dynamic pressure monitoring Air leak No load Solid load Small amount porous permeable mixed load...

Page 6: ...ments 9 2 4 1 Electrical connections 9 2 4 2 Distilled water connection 9 2 4 3 Wastewater connection 10 2 5 Installation service list 10 2 6 Handling instructions 10 3 First time operation 11 3 1 Sta...

Page 7: ...erpretation 26 8 Small steam sterilizr maintenance 27 8 1 Cleaning of items before sterilization 27 8 2 Maintenance 27 8 2 1 Cleaning and maintenance 27 8 2 2 Regular maintenance 29 9 Failure and comm...

Page 8: ...10 Tray 3 Aluminum tray with holes at the bottom for placing sterilized items Tray handling 1 Tool to take out the tray Chamber rack 1 Rack to hold the tray Fuse 2 220V 16A fuse Air filter 1 Air filt...

Page 9: ...2 Fig 1 1 front view 1 2 2 Side view Fig 1 2 side view Door handle Power switch Display...

Page 10: ...e View Fig 1 5 inside view Wastewater drainage interface A empty Power Switch Printer interface Test interface USB interface Emergency door opening device Air filter Power cable Safety valve Wastewate...

Page 11: ...475mm Capacity of Sterilizer 18 L 23 L Pressure vessel weight 4 8 kg 6 2 kg Equipment weight 59 6 kg 71 6 kg 64 8 kg 66 8 kg Built in clean water tank volume 3 8L Built in waste water tank volume 4 5...

Page 12: ...te 2 Anti electromagnetic interference is the ability of the sterilizer to work properly in the presence of electromagnetic interference 1 4 Loading method The tray rack is included in the delivery of...

Page 13: ...grasp the white plastic disc inside the protective cover gently pull distance is about 1CM At the same time use your left hand to lift the door handle up to open the door 1 7 Equipment Workflow 1 7 1...

Page 14: ...e of the vacuum test is to test the leak rate of the sterilizer No sterilization of the items in the chamber during the process Please run the vacuum test in a cold and dry chamber without loads Proce...

Page 15: ...explosion Countertop must be able to support the weight of this product Q50B 59 6 kg Q51B 61 6 kg Q70B 64 8 kg Q71B 66 8 kg Each surface of the device except the bottom must have a gap of 10 cm outwar...

Page 16: ...ted that the distilled water added should not exceed the height of the water deflector generally stored in a standard 12L distilled water tank Full tank of distilled water can complete up to 20 steril...

Page 17: ...ice are not bent or knotted 2 5 Installation service list After installation please carefully fill out the installation service form in the manual which is useful information for maintenance 2 6 Handl...

Page 18: ...y out pressure test After the pressure verification is successful the initial setting is completed The date and time can also be set as required See Chapter 6 system settings in this manual Steps of p...

Page 19: ...e may increase Don t slam the door at any time Failure to comply with the above requirements may affect the function and reliability of the door lock device Open the door only when loading and unloadi...

Page 20: ...to be sterilized Proper loading is a key factor for effective sterilization and drying You must also pay attention to the following when loading The load must be sterilized by the manufacturer if ste...

Page 21: ...hollow instruments Quick B 134 3 1 99 Wrapped unwrapped solid instruments Fabric 121 15 1 99 Fabric Select the sterilization program as follows Fig 4 1 Select sterilization program 4 3 Auxiliary opti...

Page 22: ...cotton fabrics See section 6 6 for details 4 4 Running a program 4 4 1 Start sterilization After selecting sterilization program pleaes press to start up program Sterilizer will ensure that the door i...

Page 23: ...opping the sterilization process at an earlier stage may cause an infection If the sterilization process is stopped before the start of drying the loaded items are not sterilized The use of these item...

Page 24: ...to remove the loaded items Balance when removing the tray to prevent the contents of the tray from slipping Damaged packaging devices or items may not be effectively sterilized and exposure to these...

Page 25: ...r operating cycle parameters should be fully recorded and stored for 3 years 4 7 Sterile Storage Sterile package storage time after sterilization varies with packaging material and storage conditions...

Page 26: ...d the chamber is dry In order to successfully complete the vacuum test please check that the door of sterilizer is closed and the cooling water valve is open before starting the program 1 Sterilizatio...

Page 27: ...at the system s leak rate meets the design requirements Print the information as follows PROGRAM PROPERLY EXECUTED Vacuum Test Success Leakage rate 0 13Kpa min Leak rate 0 13Kpa min Your system is ste...

Page 28: ...rainage program before loading the item for sterilization Especially when the program is abnormally stopped this situation is easy to occur Please open the door for inspection Please carry out the dra...

Page 29: ...ct date setting Press to enter date setting Time setting Date setting Language setting Warm up setting Press and to set up date Press to switch between year month day Press to exit setting Time settin...

Page 30: ...ess to select printer setting Press to enter printer setting system Date setting Language setting Warm up setting Printer setting Press and to setup printer Press to exit printer setting after complet...

Page 31: ...rying setting System reset Press and to enter into additional drying setting Press to exit additional drying setting after completion Additional drying setting Additional drying close Completion Press...

Page 32: ...Sterilization records are named after serial numbers such as 0001 TXT can be exported through the USB interface or printed by a printer Press and to select whether to perform a system reset Press to...

Page 33: ...the export completion prompt Please remove the USB flash drive Fig 7 1 Record Export 7 2 Log printing When the automatic printing function is turned on and the sterilizer is properly connected to the...

Page 34: ...Realse 0 10 105 6 0 18 27 Dry 0 10105 60 18 27 Dry End 0 9190 3 0 22 27 Ventile 0 1492 3 0 22 39 Finish 0 1492 3 0 22 39 PROGRAM PROPERLY EXECUTED Temperature 134 7 0 2 0 4 C Pressure 2 14 0 02 0 03ba...

Page 35: ...tenance operation Replace the seal Sterilize 1000 times Open the door and check whether the seal is deformed or wrinkle etc If it is deformed or wrinkled it is recommended to replace it Seal maintenan...

Page 36: ...the small steam sterilizer every year to ensure good equipment working condition and sterilization effect Maintenance and take care of this device must be performed by a professional authorized by th...

Page 37: ...s placed correctly in the chamber do not touch the wall of the pressure chamber directly and if the standard trays and tray rack are used for loading Activate automatic warm up Open the left side cove...

Page 38: ...lly disconnects the preheating system power supply circuit when the temperature exceeds the limitation of protection switch threshold temperature to realize the protection function of the preheating s...

Page 39: ...32 Piping system schematic Appendix 3 Electrical schematic diagram...

Page 40: ...Small steam sterilizer electrical schematic diagram...