6.11 RESET CONFIGURE
You can simultaneously reset most stored configuration
settings -- except those that define sensor calibration -- to
factory-set defaults (see Table D). The excluded settings
are:
•
SELECT SENSOR
•
SET SLOPE
•
TEE SIZE
•
SET OFFSET
•
TEE MATERIAL
•
OFFSET BY
•
SET PIPE ID
•
SET TO
1.
With the
top-level menu screen dis-
played, use
Ø
Ø
key
to select the “RESET CONFIGURE”
line.
2.
Press
ENTER key
to display the “RESET CONFIGURE:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (If you want to abort this
action, press
ESC key
now.)
3.
Press
ENTER key
to reset most stored configuration
settings to factory defaults (“RESET CONFIGURE:
DONE” screen appears, acknowledging reset has oc-
curred).
4.
To return to the TEST/MAINT top-level menu screen,
press
ESC key
or
ENTER key
.