PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
Rev. 1-201
46
☞
GLI Pipe Mount Sensor Setup (SET PIPE ID)
1.
Determine the actual inside pipe diameter. Actually
measure it or refer to an appropriate reference table.
NOTE:
Pipe IDs for commonly-used Schedule 40 and
Schedule 80 pipes are listed in Tables A and B
below.
Table A -- SCHEDULE 40 PIPE
Table B -- SCHEDULE 80 PIPE
Nominal
Pipe Size
Pipe ID
(inches)
Nominal
Pipe Size
Pipe ID
(inches)
3 inch
3.068
3 inch
2.900
4 inch
4.026
4 inch
3.826
5 inch
5.047
5 inch
4.813
6 inch
6.065
6 inch
5.761
8 inch
7.981
8 inch
7.625
10 inch
10.020
10 inch
9.562
12 inch
11.938
12 inch
11.374
14 inch
13.124
14 inch
12.500
16 inch
15.000
16 inch
14.312
18 inch
16.876
18 inch
16.124
20 inch
18.812
20 inch
17.938
24 inch
22.624
24 inch
21.562
2.
With the
submenu screen displayed
and the “SET PIPE ID” line selected, press
ENTER key
to display a screen like
. Use
arrow
keys
to adjust the displayed value to exactly match the
actual pipe ID (inside diameter), and press
ENTER key
to enter the value.
This completes pipe-mount sensor data entry/calibration.