PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
Rev. 1-201
54
SET 0/4 mA and
20 mA VALUES
(range expand)
2.
Press
ENTER key
to display
.
3.
With the “SET OUTPUT 1” line selected, press
ENTER
key
to display
.
4.
With the “SET PARAMETER” line selected, press
ENTER key
to display a screen like
.
Use
Ø
Ø
and
×
×
keys
to view the choices. With the de-
sired choice displayed, press
ENTER key
to enter this
selection.
You can set parameter (or calculated measurement) values
to define the endpoints at which the minimum and maximum
output values are desired.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “SET 4 mA VALUE” line.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to set the displayed
value at which 0/4 mA is desired, and press
ENTER
key
to enter the value.
3.
After the
screen re-appears, use
Ø
Ø
key
to select the “SET 20 mA VALUE” line.