PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Rev. 1-201
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
77
6.10 SIM Setting
☞
☞
After selecting the type of simulated measurement (sub-
section 6.9), set the desired simulation value.
1.
With the
top-level menu screen dis-
played, use
Ø
Ø
key
to select the line below the “SELECT
SIM” line.
NOTE:
The line directly below the “SELECT SIM” line
always relates to and shows the type of simu-
lated value choice selected in subsection 6.9.
For example, if the choice was “VOLUME A,”
this line becomes “SIM VOLUME A.”
2.
Press
ENTER key
to display a screen like
.
NOTE:
Both analog output signals are now active.
They have a mA value that corresponds to the
measurement value shown on this screen. (The
relays, depending on their configured settings,
may also respond to this simulation value.)
3.
Use
arrow keys
to adjust the displayed simulation
value to the desired value.
4.
To end the simulation and return to the TEST/MAINT
top-level menu screen, press
ESC key
or
ENTER key
.