User Manual for Vital Signs Monitor
29
4.7.3 NIBP Setup
Figure 4.13A NIBP Setup
Figure 4.13B Tourniquet Setup
Figure 4.13C Adjust Setup
NIBP Setup Screen Description:
“SYS
”:
systolic pressure alarm switch; “
” indicates systolic pressure alarm is on; “
”
indicates systolic pressure alarm is off.
“SYS Hi”:
high limit of systolic pressure alarm; range: “32~250” mmHg.
“SYS Lo”:
low limit of systolic pressure alarm; range: “30~248” mmHg.
“DIA
”:
diastolic pressure alarm switch; “
” indicates diastolic pressure alarm is on; “
”
indicates systolic pressure alarm is off.
“DIA Hi”:
high limit of diastolic pressure alarm; range: “22~230” mmHg.
“DIA Lo”:
low limit of diastolic pressure alarm; range: “20~228” mmHg.
“MAP
”:
mean arterial pressure alarm switch; “
” indicates mean arterial pressure alarm is on;
“
” indicates mean arterial pressure alarm is off.
“MAP Hi”:
high limit of mean arterial pressure alarm; range: “28~242” mmHg.
“MAP Lo”:
low limit of mean arterial pressure alarm; range: “26~240” mmHg.
“Mode”:
NIBP measuring mode, “manual”, “AUTO 1”, “AUTO 2” …“AUTO 240” and “STAT”
etc. options. “AUTO 1” means NIBP measurement takes once every one minute automatically;
“AUTO 60” means NIBP measurement takes once every 60 minutes automatically; In AUTO mode,