Chapter 11 Appendix
11.1 Prompt information explanations
Mute C-D: XXX seconds
Alarm silence count down: XXX seconds
NIBP C-D: XXX seconds
NIBP auto measuring cycle count down: XXX seconds
TOUR C-D: XXX seconds
Tourniquet alert count down: XXX seconds
Probe off
SpO
2
probe fells off
PR over limit
PR value exceeds the high/low alarm limit
SpO
2
over limit
SpO
2
value exceeds the high/low alarm limit
SYS over limit
Systolic pressure value exceeds the high/low alarm limit
DIA over limit
Diastolic pressure value exceeds the high/low alarm limit
MAP over limit
MAP value exceeds the high/low alarm limit
NIBP error 1#
Sensor or other hardware error
NIBP error 2#
Very weak signal because of the cuff, or the patient has very weak pulse
NIBP error 3#
Blood pressure amplifier overflow due to excessive movement
NIBP error 4#
Leaking during the pneumatic device testing
Cuff error
Cuff is not wrapped correctly, or is not connected
NIBP error 5#
Abnormal condition of CPU, such as register overflow, divided by zero
Air leak
Air moving part, tube or the cuff leak air
NIBP over range
The measurement range exceeds 255mmHg (for neonates: over 135 mmHg)
Over motion
The repeated measurement due to moving, excessive noise during the
stepping inflation and measuring pressure and pulse, e.g. during patient
shaking motion
Over pressure
Cuff press exceeds the safety limit value of software. Limit value for adult:
290mmHg; Limit value for pediatric: 145mmHg;
Or caused by cuff extrusion or flapping cuff with force.
NIBP timeout
Adult measurement is more than 120 seconds, neonate measurement is more
than 90 seconds.