User Manual for Vital Signs Monitor
25
4.5 SpO
2
List Screen
Short pressing “
Display” key to shift the screen to SpO
2
List screen, as shown in Figure 4.7
Figure 4.7 SpO
2
List
The first column is the date, the second column is SpO
2
measuring time, the third column is SpO
2
value, and
the fourth column is pulse rate. Up to 2000 groups data which are lately measured can be stored in the
monitor when it is out of power..
4.5.1 Operation Instructions
On SpO
2
List screen, if SpO
2
value measurement is more than 8 groups, press “
▲
” key or “
▼
” key to turn to
previous/next page for view other measurement values. If SpO
2
measurement is not more than 8 groups, the
keys “
▲
”or “
▼
” are not effective.
“
” key:
short press this key to shift to next display view. Long press to enter the screen of
Empty history records, the user can delete all SpO
2
records according to prompt.
“
” key: print SpO
2
list if the monitor equips with built-in printer, press this key again to stop
printing
“
”
key:
start/cancel measuring NIBP.
“
” key:
alarm silence switch; press it to enable/disable alarm silence.
4.6 ECG Recall Screen
Short pressing “
Display” key to shift the screen to Default ECG Recall screen, as shown in Figure 4.8.
On this screen, the latest 120-hour ECG data stored in monitor can be reviewed.
Figure 4.8 Default ECG Recall Screen