User Manual for Vital Signs Monitor
31
calibration the “Stop” shifts to “Start”)
NIBP Cali Mode 2
:
Receiving the exterior pressure. The exterior pressure source pressurize to the
module to proceed the pressure calibration. Move the cursor to NIBP calibration mode 2”Start” button,
click the OK button to begin the NIBP calibration.(Meanwhile, the “Start” shifts to “Stop”, after the
calibration the “Stop” shifts to “ Start”)
Gas leak
: Move the cursor to Gas leak ”Start” button, click the OK button, the pump inflates to
certain pressure and then the valve will be closed for leak detection for ten seconds, then the blood
pressure module will deflate automatically and the screen displays measurements.
"
The NIBP calibration and Gas leak detection can only be carried on when the NIBP measurement is
set to mode “Manual”.
"
Other buttons are disabled except “
■
” OK button and “
” Power button during NIBP calibration
and Gas leak detection.
"
Make sure the “
■
” OK button is off after the test, or the user could not do other operations.
NIBP Mode Setup Shortcut Screen Descriptions:
In waveform display screen or trend graphic screen or NIBP list screen longtime press “
”key about
3 seconds can enter into the screen shown in Figure 4.13D. Please refer to “NIBP Setup Screen
Description” for more detailed information.
Figure 4.13D NIBP Mode Setup Shortcut Screen