User Manual for Vital Signs Monitor
16
3.3.4 TEMP Transducer Connection
Connecting methods:
1. Attach the transducers to the patient firmly;
2. Connect the cable to TEMP probe connector in the front panel.
Note: When unplugging the probe, be sure to hold the head of the connector and pull it out.
3.3.5 Loading printing paper
Operation procedures for loading printing paper:
1. Press both “OPEN” notches with force on printer shield with two thumbs to open it.
2. Move the tab of rubber roller lock at the left 90°upwards to unlock it, refer to the following
figure with mark
①
.
3. Cut one end of the paper into triangle, and load the paper from the underside of the rubber roller.
4. Turn the roller clockwise to get the paper rolled, and put the paper roll into the compartment.
5. Pull the paper out of paper slot on the shield.
6. Move the tab of the rubber roller lock 90° downwards to lock it.
7. Put the shield back in position and secure it.
Operation procedures for taking out printing paper roll:
1~2 steps are the same with the 1~2 steps mentioned above for loading printing paper.
3. Roll the loading roller anti-clockwise and pull the paper out.
4~5 steps are the same with the 6~7 steps mentioned above for loading printing paper.
Figure 3.8 Loading and taking out printing paper
P8 printer may be used due to the different configuration.
P8 printer operation instruction:
Power indicator: green light shows the power is on, while the monitor is out of power, the green light is off.
Error indicator: red light is constant which shows the printer is out of paper, or the printing paper does not
install well. When the printer installs normally, the red light is off.