11.2 Default Alarming Values and Setup Range
The default alarming value:
Mode
Parameter
Adult
Pediatric
Neonate
HR
High limit
180bpm
200bpm
220bpm
Low limit
40bpm
50bpm
50bpm
SYS
High limit
180mmHg
130mmHg
110mmHg
Low limit
60mmHg
50mmHg
50mmHg
DIA
High limit
120mmHg
90mmHg
90mmHg
Low limit
50mmHg
40mmHg
30mmHg
MAP
High limit
160mmHg
110mmHg
100mmHg
Low limit
50mmHg
40mmHg
30mmHg
SpO
2
High limit
100%
100%
100%
Low limit
90%
85%
85%
Pulse rate
High limit
180bpm
200bpm
220bpm
Low limit
40bpm
50bpm
50bpm
TEMP
High limit
39.0
℃
39.0
℃
39.0
℃
Low limit
35.0
℃
35.0
℃
35.0
℃
The high and low limits setting range:
Mode
Parameter
Adult
Pediatric
Neonate
HR
High limit
(1~350) bpm
(1~350) bpm
(1~350) bpm
Low limit
(0~349) bpm
(0~349) bpm
(0~349) bpm
SYS
High limit
(30~280) mmHg
(30~200) mmHg
(30~135) mmHg
Low limit
(29~279) mmHg
(29~199) mmHg
(29~134) mmHg
DIA
High limit
(11~232) mmHg
(11~150) mmHg
(11~100) mmHg
Low limit
(10~231) mmHg
(10~149) mmHg
(10~99)mmHg
MAP
High limit
(21~242) mmHg
(21~165) mmHg
(21~110) mmHg
Low limit
(20~241) mmHg
(20~164) mmHg
(20~109)mmHg
SpO
2
High limit
1~100%
1~100%
1~100%
Low limit
0~99%
0~99%
0~99%
Pulse rate
High limit
(1~300) bpm
(1~350) bpm
(1~350) bpm
Low limit
(0~299) bpm
(0~349) bpm
(0~349) bpm
TEMP
High limit
(0.1~60)
℃
(0.1~60)
℃
(0.1~60)
℃
Low limit
(0~59.9)
℃
(0~59.9)
℃
(0~59.9)
℃