User Manual for Vital Signs Monitor
27
4. Press “
” to exist from Setup Menu Screen.
☞
At Setup Menu Screen or its submenu screen, when pressing “
” key, the default display screen
will be printed.
The following will cover each functional parameter’s setting up.
Note
: If you disable Hi and Lo limit alarm function of parameter monitoring, all the alarms related to
its parameter monitoring will be disabled as well.
4.7.1 ECG and Temperature Setup
Figure 4.11 ECG/TEMP Setup Screen
Screen Description:
“Lead”:
ECG lead selection:
Ⅰ
,
Ⅱ
,
Ⅲ
, AVR, AVL, AVF or
V
.
“Gain”:
ECG waveform scale:
“×1/4”-
Waveform scaled with 1/4 of the base gain
“×1/2”-
Waveform scaled with half of the base gain
“×1”-
Waveform scaled with base gain
“×2”-
Waveform scaled with twice of the base gain
“1mV”:
generating internal 1mV calibration signal. This signal is used to test the function of the
machine. It is not used during normal operation. The default set is off.
“Enhancement mode”:
the switch of ECG enhancement mode.
“HR
”:
HR alarm switch; “
” indicates HR alarm is on; “
” indicates HR alarm is off.
“HR Hi/Lo”:
high/low limit of HR alarm.
“TEMP
”:
temperature alarm switch; “
” indicates temperature alarm is on; “
” indicates
temperature alarm is off.
“TEMP Hi/Lo”:
high/low limit of temperature alarm.
“Unit”:
body temperature unit. Two options: “
℃
” or “
℉
”. Conversion relation: 1
℉
= (
℃
X1.8) +32.