User Manual for Vital Signs Monitor
32
4.7.4 Nurse Call
Figure 4.14 Nurse Call Setup Screen
Screen Description:
“Output level”:
two options “low” or “high” output levels are available.
When the calling system in hospital works in “Normal Open” mode, “low level” should be selected.
When the calling system in hospital works in “Normal Close” mode, “high level” should be selected
“Source”
:
three kinds of alarm sources can trig the nurse call: high level alarm, medium level alarm
and low level alarm (multi-optional). If you don’t make choice, nurse call signal will not be sent out.
“Duration”:
two options “pulse” or “continuous” output modes are available;
“continuous”:
the continuous mode of output means the nurse call signal will keep until the selected
alarm source(es) disappear, i.e. the signal will last from starting alarm to stopping alarm.
“pulse”:
the output nurse call signal is pulse signal which lasts for 1 second. When several alarms
occur at the same time, only one pulse signal will be sent out.
Note:
Nurse Call function can not be regarded as main alarm notice method, please do not
entirely relay on it. You should combine parameter values with alarm level and patient’s clinical
behavior and symptom to determine patient’s status.