“SOFT” COAGULATION AT LOW VOLTAGE – Low superficial sparkling effect, and good deep
coagulating effect
This current guarantees a good deep coagulating effect, and a very low superficial coagulating
effect. It is suitable for coagulations performed or by forceps/surgical instrument or by
touching the tissues with an active ball electrode of 4-5mm at least.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
SOFT
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (acute acoustic signal, and blue Led), press on the blue
pedal of the twin foot-switch.
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 30-40W.
“SPRAY” COAGULATION AT VERY HIGH VOLTAGE – Very strong superficial sparkling effect, and
good deep coagulating effect
This current guarantees a good deep coagulating effect, and a very strong superficial
coagulating effect. It is suitable both for the coagulations performed with forceps/surgical
instrument, and for coagulations performed directly with the active electrode, even far away
from the tissues.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
SPRAY
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (acute acoustic signal, and blue Led), press on the blue
pedal of the twin foot-switch.
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 20-30W.
ADVICE
For the Cut, use as follows:
Current
PURE
, for a cut without coagulating effect and constant delivery of the power (from 20-30W for papillotomy
or polipectomy with polypus
until 5mm, and from 30-40W with polypus
6mm or more);
Current
BLEND-1
, for a cut with a soft coagulating effect and constant delivery of the power (from 20-30W for
papillotomy or polipectomy with polypus
until 5mm, and from 30-40W with polypus
6mm or more);
Current
ENDO,
for a cut-coagulation combination (alternated phases of cut and coagulation), and self regulation of
the power (from 40-50W for papillotomy or polipectomy with polypus
until 5mm, and from 70-80W with polypus
6mm or more).
For the Coagulation, use as follows:
Current
FULG,
from 30-40W;
Current
PIN POINT
, from 40-50W, for a superficial effect which is softer than the one of the
FULG
coagulation.