Use the flexible electrodes of the endoscopes, from 20-30W for the papillotomy, from 20-30W for the
polipectomy with polypus
5mm, and from 40-50W for polypus
6mm or more.
COAGULATING CUT (BLEND-2)
–
Not suitable at all.
COMBINATED CUT (CUT ALTERNATED TO COAGULATION) (ENDO)
– For flexible endoscopy with
cutting phases alternated to coagulation phases
This is a constant voltage current and it is controlled by an automatic power self adjustment
system (
APC System
) according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key, select the current
ENDO
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal
of the twin foot-switch.
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 70-80W for the papillotomy, from 70-80W for the
polipectomy with polypus
5mm, and from 90-100W for polypus
6mm or more.
CURRENTS FOR COAGULATION
“FULGURATION” COAGULATION AT HIGH VOLTAGE - Strong superficial sparkling effect, and
optimum deep coagulating effect
This current guarantees a strong coagulating effect, both deep and superficial, and therefore it
is suitable both for coagulations performed with forceps/surgical instrument, and for
coagulations performed by grazing the tissues with the active electrode.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
FULG
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (acute acoustic signal, and blue Led), press on the blue
pedal of the twin foot-switch.
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 20-30W.
“PIN POINT” COAGULATION AT MEDIUM VOLTAGE – Medium-Low superficial sparkling effect, and
optimum deep coagulating effect
This current guarantees a strong deep coagulating effect, and a normal superficial coagulating
effect. It is suitable both for coagulations performed with forceps/surgical instrument, and for
coagulations performed by grazing the tissues with the active electrode, just in case the
operators prefer a superficial effect which is less strong than the one of the FULG coagulation.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
PIN POINT
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (acute acoustic signal, and blue Led), press on the blue
pedal of the twin foot-switch.
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 20-30W.