Another active electrode, for the normal electro-surgery by hand-switches (connected to the socket
4-B
)
pressing the
yellow
keyl for the pure/coagulating cut and the
blue
key for the coagulation currents.(See Par.
“CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”);
C. !!! BIPOLAR FUNCTIONING !!!
The bipolar functioning is always available. See Par. “BIPOLAR MODE”.
PRE-SET MEMORIES (nr.6 and 7)
–
These two programs allow to use, always by the same twin foot-
switch, two different monopolar active electrodes (one blade or needle electrode for cutting inserted in an
handle and one forceps for coagulation with connection cable to the unit) simply setting the same
currents or output powers and alternating the two programs (i.e. the program 89 or for using the cutting
or coagulating current “instrument for cutting or coagulating cutting” and the program 90 for using the
coagulating current “instrument for coagulation”)
The pre-setting does not concern the currents or their powers which have to be set according to the operators
needs and memorised as specified in the first part of the paragraph (for this kind of use, it is advisable to select and
set the same currents and the same powers in both memories).
The pre-setting of these memories allows to use the unit as follows:
A.
Selecting the memory 89,
one electrodes-holder handle
(connected to the socket
5-A
) can be used by the
foot-switch pedal. (See Par.
“CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”) pressing (on the
twin foot switch) the
yellow
pedal for the pure/coagulating cut and the
blue
pedal for the coagulation currents;
B.
Selecting the memory 90,
one coagulation forceps
(connected to the socket
6
)
can be used by the foot-switch
pedal
. (See Par. “CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”) pressing (on the twin foot
switch) the
blue
pedal for the coagulation currents;
The activation modes and the outputs available can not be changed, but it is possible to set pre-set
currents/output powers and memorize the settings into these programs by pressing the key MEM of the
area B (See the first part of this paragraph).
PRE-SET MEMORIES (nr.8)
– Use in flexible endoscopic surgery or in endoscopic surgery
under liquid.
The pre-setting concerns:
1.
Activation mode and output available.
2.
In these programs are not set output powers or currents because this choice depends on the operator.
The activation modes and the outputs available can not be changed, but it is possible to set standard
currents/output powers and memorize the settings into these programs by pressing the key MEM of the
area B (See the first part of this paragraph).
The pre-setting of these memories allows to use the unit as follows:
A.
With the flexible endoscope or with the rigid resectoscope in(connected to the socket
5-A
) by pedal foot-
switches (See Par. “CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”), pressing (on the twin foot
switch) the
yellow
pedal for the pure/coagulating cut and the
blue
pedal for the coagulation currents;
B.
!!! BIPOLAR FUNCTIONING !!!
The bipolar functioning is always available even if it is not used in the endoscopy under liquid but only, hardly
ever, in flexible
endoscopy. See Par. “BIPOLAR MODE”.
PRE-SET MEMORIES (nr. 9)
– Use in bipolar mode only.
The pre-setting does not concern the current types that have to be used, nor their powers, which have to be set
according to the operators’ needs, and therefore memorised as shown at points 3 and 4 of the previous paragraph.
The pre-setting of these memories allows to use the unit as follows:
A.
Without connecting the neutral electrode, as it is not required (through the alarm Led, the safety circuit just
informs the operators that the neutral electrode is not connected);
B.
Without using the monopolar currents (their displays are switched off);
It is possible to set standard programs including: the currents and their powers, the number of twin foot-
switches (1 or 2) and memorize the settings into the programs by pressing the key MEM of the area B (See
the first part of this paragraph).
SELF-DIAGNOSIS and SELF TEST SYSTEM, CONTROL of the ERROR CODES
The unit is equipped with a self-diagnosis system, which automatically intervenes in case of functioning problems,
included the delivery of powers if are higher than the set ones. It stops the device and informs the operators
through specific alarm signals (
Error Codes
), which can be either acoustic (loud and intermittent sound) or also
visual (see the Tables
Codes for Wrong Use or Activation Circuit Fault
–
Codes for System Fault
).