Use cutting electrodes (blade, needle, Hook type, or the external part of the dissector, in laparoscopy) from 30-
40W.
COMBINATED CUT (CUT ALTERNATED TO COAGULATION) (ENDO)
– For flexible endoscopy with
cutting phases alternated to coagulation phases
This is a constant voltage current and it is controlled by an automatic power self
adjustment system (
APC System
) according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key, select the current
ENDO
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the foot-switch, or the
yellow
key of the hand-switch handle.
Instructions for use, adjustments and electrodes
Ideal only for flexible endoscopic surgery (i.e. polipectomy or papillotomy).
CURRENTS FOR COAGULATION
“FULGURATION” COAGULATION– Strong superficial sparkling effect, and optimum deep coagulating
effect
This current guarantees a strong coagulating effect, both deep and superficial, and therefore it
is suitable both for coagulations performed with forceps/surgical instrument, and for
coagulations performed by grazing the tissues with the active electrode.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
FULG
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
acute
acoustic signal, and
blue
Led), press on the
blue
pedal of the twin foot-switch , or the
blue
key of the hand-switch electrodes-holder handle. It is also possible to deliver the current by simply closing
the coagulation forceps, just in case a hand-switch forceps.
Instructions for use, adjustments and electrodes
It can be used with all types of electrodes (ball, blade, or needle type, insulated coagulation forceps, loop
electrodes in conization,
Hook electrodes, dissectors, and forceps in laparoscopy) from 40-50W.
“PIN POINT” COAGULATION – Medium-Low superficial sparkling effect, and optimum deep
coagulating effect
This current guarantees a strong deep coagulating effect and a normal superficial coagulating
effect. It is suitable both for coagulations performed with forceps/surgical instrument and for
coagulations performed directly with the active electrode, just in case the operators prefer a
superficial effect which is less strong than the one of the FULG coagulation.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
PIN POINT
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
acute
acoustic signal, and
blue
Led), press on the
blue
pedal of the twin foot-switch, or the
blue
key of the hand-switch electrodes-holder handle. It is also possible to deliver the current by simply closing
the coagulation forceps, just in case a hand-switch forceps is used.
Instructions for use, adjustments and electrodes
It can be used with all types of electrodes (ball, blade, or needle type, insulated coagulation forceps, loop
electrodes in conization,
Hook electrodes, dissectors, and forceps in laparoscopy) from 40-50W.