Error description
Error code
Stored code in
E
2
PROM (select “0d”
in CAL mode)
1
I
2
CBUS connection lost (during the use)
6 slow sound
repeatitions; after
fast repeatitions
43
Master-Slave connection lost (during selftest phase)
7 slow sound
repeatitions; after
fast repeatitions
41
Slave microcontroller failure
8 slow sound
repeatitions; after
fast repeatitions
44
E
RRORS RELATED
S
LAVE MICROCONTROLLER
Failure in a Slave microcontroller internal peripheral
Err 60
3C
Slave microcontroller connection failed
Err 68
44
Errors related to HF power section
Failure in internal dummy load used for selftest
Err 39
27
Failure in HF output current measurement (lower than expected)
Err 51
33
Failure in HF output current measurement (higher than expected)
Err 52
34
Failure in bipolar output power( lower than expected)
Err 43
2b
Failure in bipolar output power( higher than expected)
Err 42
2A
Failure in monopolar circuit when PURE CUT mode is used
Err 40
28
Failure in monopolar circuit when SPRAY COAG mode is used
Err 41
29
Failed modulation signal for monopolar BLEND-1 mode
Err 44
2C
Failed modulation signal for monopolar FULG mode
Err 45
2d
Failed modulation signal for monopolar SPRAY mode
Err 46
2E
Output peak voltage reading failed (lower than expected)
Err 47
2F
Output peak voltage reading failed (higher than expected)
Err 48
30
Output peak current reading failed (lower than expected)
Err 49
31
Output peak current reading failed (higher than expected)
Err 50
32
HF output power higher than expected
Err 98
62
E
RRORS RELATED TO THERMAL PROTECTION
Power supply thermal protection
Err 27
1b
HF power section thermal protection
Err 28
1C
ENVIRONMENTAL and ATMOSPHERIC CONDITIONS for USE, TRANSPORT
and STORAGE
The environmental use and preservation conditions are the following ones:
Temperature (°C): +10 ÷ +40
– Humidity: 30% ÷ 75% - Pressure (hPa): 700 ÷ 1060.
The unit must not be used at less than 30cm from the wall or other objects that can obstruct the
ventilation areas, and it must be placed on a trolley or a support (there are two screwed points under the
basis for the fixing).
When the unit is not used, it must be kept in a dry place, not dusty. Be aware that no liquid is poured on
it.
The environmental transport and storage conditions are the following ones:
Temperature (°C): -40 ÷ +70 - Humidity: 10% ÷ 95% - Pressure (hPa): 500 ÷ 1060.
For the shipment of the unit, we suggest to use the original packaging, or at least a new one that can guarantee the
same reliability.
MAINTENANCE and DISPOSAL
The unit must be regularly checked (once per year) by qualified personnel, even better if by the
Manufacturer himself. Check always the perfect condition of the accessories, because
otherwise their use can be dangerous.