FUNCTIONING FOR THE UNDER LIQUID ENDOSCOPIC SURGERY
(TUR, TURP, TURV, etc.)
In order to favour this kind of use, the unit is equipped with 1 specific pre-
set program (See Par. “PROGRAMS AND
MEMORIES”, programs 8), where the following features are already memorized:
1.
Activation mode and output available for the connection of the resectoscope
2.
In this program are not set output powers or currents because this choice depends on the operator.
In all cases, it is possible to set standard pre-set currents/output powers and therefore memorize these
settings into the different programs (See Par.
“PROGRAMS AND MEMORIES”).
CONNECTION AND USE OF THE ACCESSORIES
A) Socket (5-A, the led A under is lighted on): monopolar rigid resectoscope
with pedal foot-switch
(See Par. “CONNECTION AND USE OF THE
PEDAL FOOT-
SWITCHES”).
Use only this socket to connect the above mentioned instruments. The use of different sockets provokes a
damage to the unit.
Just in case, ask for:
The adaptors for the cables with different plug:
Adaptor
-
34-9590
, for cables with non insulated plugs
from 2 to 8mm, or with insulated plug
4mm;
B) Socket (7- NEUTRAL PLATE): neutral electrode.
According to the connection possibilities listed above, the unit allows to:
A. Use the resectoscope by foot-switch activation
(See PAR.
“CONNECTION AND USE OF THE PEDAL
FOOT-
SWITCHES”
).
Just press on the
yellow
pedal for the pure cutting/coagulating cutting currents, and on the
blue
pedal for the
coagulation currents.
B. !!! BIPOLAR FUNCTIONING!!!
The bipolar functioning is always possible. See Par. “BIPOLAR MODE”.
MONOPOLAR CURRENTS, ELECTRODES, ADJUSTMENT OF THE POWERS, ADVICE
CURRENTS FOR CUT
PURE CUT (PURE)
– Cut without coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key
MODE
, select the current
PURE
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the twin foot-switch.
Instructions for use, adjustments, and electrodes
Use the specific electrodes for the resectoscope from:
70-80W with wire loop electrodes, for soft tissues (i.e. into the bladder);
90-100W with wire loop electrodes, for strong tissues (i.e. TUR);
160-180W with ribbon loop (cutting/vaporization) electrodes, for soft tissues (i.e. into the bladder);
200-210W with ribbon loop (cutting/vaporization) electrodes, for strong tissues (i.e. TUR);
100-120W with small rolls for vaporisation;
200-210W with big rolls for vaporisation.
COAGULATING CUT (BLEND-1)
– Cut with soft coagulating effect