background image

 

 

 
 

 

 

 

Dateiname mit Revisionsstand: 

erstellt am: 

erstellt von: 

geprüft von: 

freigegeben von: 

Seite 

ESU-X 300NT_EN B 

18.04.2018 

ssta 

cgro 

cgro 

5 von 35 

 
 
 

and  a  decrease  of  the  output  power  into  the  application  point  (which  leads  the  operator  to  raise  it, 
dangerously);    

Never exceed 1/3 of the max. output power for each monopolar current when using the paediatric neutral 
electrodes, or 1/5 when using the electrodes for babies;  

Use the disposable neutral electrodes only once, by paying attention to the instructions on the packaging. 
Make  sure  they  have  the  right  dimensions  (standard  for  adults  with  “weight  body  higher  than  15  Kg”: 
approx. 136cm

2

; standard for children with “weight body from 5 to 15 Kg”: approx. 84cm

2

);

 

 

As the space between the neutral electrode and the operating area represents a sort of “path” for the HF 
current,  be  sure  that  it  is  not  diagonal  as  regards  the  body,  nor  on  the  heart.  Remember  also  that  the 
metallic  elements  (prosthesis,  catheters,  etc.)  on  the  path  of  the  current  may  cause  accumulations  of 
current with consequent heating/burns of the surrounding tissue;     

12.  Position the cables of the electrodes in a way that they do not touch the patient or any other conducting part. 

During the operations, place the unused active electrodes on insulating materials, far away from the patient;   

13.  Always  use  the  lowest  possible  power.  Bear  this  warning  in  mind  when  intervening  on patients  (children or 

babies) for whom small neutral electrodes are used (see also point 11); 

14.  Choose the bipolar technique, when operating on small portions of tissue or in cavities;  
15.  Try to respect as much as possible the suggested working times, and avoid useless short-circuits between the 

active electrode and the neutral one; 

16. 

Get in contact with the Technical Service for the use of the “disposable” electrodes; 

17.  When  the  device is switched on,  check  all  the settings  before  using it  on  the  patient,  and  remember  that  a 

failure can provoke an undesired increase of the power; 

18.  Remember that also the use of too low powers, if combined with some particular electrodes or accessories, 

can cause side effects: for example, when using the Argon gas, the risk of embolism raises if the power of the 
spray coagulation is not able to produce quickly a rapid and impermeable eschar on the target tissues;   

19.  The unit must not be used for final purposes other than those listed in this manual. 

 
THE HF LEAKAGE CURRENTS CONTROL CIRCUIT 

The unit is equipped with a leakage currents to earth control circuit because these currents represent one possible 
source for undesired burns on the patient or the operators (i.e. a patient who gets in touch with a metallic part of 
the operating table or with wet/damp towels, a patient who is placed on a water mattress for surgical needs, an 
operator who gets in touch with instruments or endoscopes, etc. They all are possible causes of the increase of 
such currents).  
When the leakage currents to earth overcome 150mA (limit established by the rules), the circuit intervenes as 
follows: 

-  It automatically reduces the output power, so that the leakage currents come back within the agreed limits;   
-  It gives an alarm signal to the operators (red Led 

G-

HF LEAKAGE

 

on). 

 

SAFETY CIRCUIT OF THE NEUTRAL ELECTRODE 

The neutral electrode connection control circuit (Red Led - area 

A

) operates in the following manners: 

1) 

With disposable/reusable electrodes with one single section (non split)

. The circuit controls if the neutral 

electrode is connected to the cable, and if the latter is integral and correctly connected to the unit  
 
 
(socket 

NEUTRAL PLATE-7

). If this is not the case, it stops the delivery of the power and gives a luminous 

alarm signal (the Red Led is lit, Error Code “no Np”) and a buzzer (loud, intermittent);   

 

 

2) 

With  disposable/reusable  electrodes  with  twin  section  (split)

.  The  circuit  works  as  described  above  at 

point 1, but it also checks if the quality of the contact between the electrode and the patient tissues is good 
enough. It operates as follows:  
a)  When the contact is optimum, the circuit does not intervene;  
b)  When the contact decreases to less than approx. 50% of the surface of a standard electrode for adults 

well  attached,  the  circuit  intervenes  automatically  reducing  to  max.  200W  the  output  powers  (if  higher 
levels have been selected);   

c)  When  the contact  further  decreases,  the circuit  completely  stops  the  delivery  of the power,  by  giving  a 

luminous alarm signal (The Red Led is lit, Error Code “no Np”) and also a buzzer (loud, intermittent).  

3) 

When  only  the  memories  for the  bipolar use  (it  does  not  require any  neutral  electrode)  are  selected 
(memories from 96 to 99)

,  

      

  the circuit does not intervene (the 3 Led are lit, but only to show that the electrode is not connected).   

 

INITIAL CHECKS  

Initial checks 
1. 

Make sure that the mains power supply corresponds to the technical data (see the data label on the backside 
of the unit), and connect the unit with the mains switch (

b2-green-

 on the backside of the unit) off;

 

2. 

For a possible equipotential connection, use the plug (

b4

 on the backside of the unit), and the device (

b6

 on 

the backside of the unit) to adjust the functioning acoustic signals (the max. is clockwise). The alarm signals 
cannot be regulated.     

Summary of Contents for ESU-X 300 NT

Page 1: ...stra e 2 78532 Tuttlingen Germany Tel 49 7461 96624 0 FAX 49 7461 96624 22 E Mail info geister com Read instruction for use 0123 Medical product higher class I and shortcut for Notified Body T V S D P...

Page 2: ...oagulation G Alarm Led for the HF leakage currents control circuit H Setting of the standard twin foot switch pedal functioning PEDAL 1 I Setting of the connection sockets for the bipolar electrodes l...

Page 3: ...th sufficient density the cellular liquid of the tissues it warms it and generates what follows 1 a heating which is so rapid that the vapour pressure into the cells breaks their membranes and provoke...

Page 4: ...an electrode in bad conditions or an electrode with connection defects causes a reduction of the output power as it does not have any good contact with the tissues 6 Remember that even if compliant wi...

Page 5: ...circuit because these currents represent one possible source for undesired burns on the patient or the operators i e a patient who gets in touch with a metallic part of the operating table or with wet...

Page 6: ...dal or coagulation blue pedal BIP to activate the bipolar currents of cut coagulating cut yellow pedal or coagulation blue pedal except for the coagulation AUTO which is not activated by any pedal foo...

Page 7: ...d switch handle forceps for coagulation with hand switch for coag only in alternative Socket 5 A the led A under is lighted on handle forceps for coagulation laparoscopic instrument for the use by foo...

Page 8: ...ust the power with the regulation keys UP DOWN For the delivery loud acoustic signal and yellow Led press on the yellow pedal of the twin foot switch or the yellow key of the hand switch handle Instru...

Page 9: ...the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch or the blue key of the hand switch electrodes holder handle It is also possible to deliver the current b...

Page 10: ...r the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch or the blue key of the hand switch electrodes holder handle It is also possible to deliver the current...

Page 11: ...Dateiname mit Revisionsstand erstellt am erstellt von gepr ft von freigegeben von Seite ESU X 300NT_EN B 18 04 2018 ssta cgro cgro 11 von 35...

Page 12: ...USE OF THE PEDAL FOOT SWITCHES Press on the yellow pedal for the pure coagulating cut currents and the blue pedal for the coagulation currents C BIPOLAR FUNCTIONING The bipolar functioning is always...

Page 13: ...ith the active electrode This current is controlled by an automatic system ADC System for the self regulation of the functioning which guarantees that the power selected by the operators remains const...

Page 14: ...med with forceps surgical instrument and for coagulations performed directly with the active electrode even far away from the tissues This current is controlled by an automatic system ADC System for t...

Page 15: ...ted above the unit allows to A Use the resectoscope by foot switch activation See PAR CONNECTION AND USE OF THE PEDAL FOOT SWITCHES Just press on the yellow pedal for the pure cutting coagulating cutt...

Page 16: ...keys UP DOWN For the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch Instructions for use adjustments and electrodes Use the loop ball electrodes for the r...

Page 17: ...for a cut without coagulating effect Current BLEND 1 from 90 100W on the prostate for a cut with soft coagulating effect For the Cut with ribbon loop electrodes use as follows Current PURE from 160 18...

Page 18: ...informs the operators that the neutral electrode is not connected Impossibility to use the monopolar currents the displays for the monopolar currents powers are switched off Activation modes and outp...

Page 19: ...he unit 0 5sec on the left 5sec on the right For the switching off it is possible either to disconnect the forceps from the tissues or wait until the current disconnects automatically once the tissues...

Page 20: ...thout any sticking of the tissues on the electrode the above mentioned powers maybe have to be changed through small variations of 10 15W each time For the Coagulation When bipolar forceps are used it...

Page 21: ...ion Is it possible to modify the adjustments of the power s and or the selected settings of a memory before the use Yes it is always possible through the different adjustment selection keys and this v...

Page 22: ...gon gas flow if it is used can obviously be changed by setting the desired levels and by pressing the memorisation key MEM of the area B The pre setting of these memories allows to use the unit as fol...

Page 23: ...raph PRE SET MEMORIES nr 8 Use in flexible endoscopic surgery or in endoscopic surgery under liquid The pre setting concerns 1 Activation mode and output available 2 In these programs are not set outp...

Page 24: ...o the neutral electrode broken cable non connected cable to the unit or to the neutral electrode neutral electrode split type not attached to the tissues no nP Failure into the neutral electrode contr...

Page 25: ...nopolar SPRAY mode Err 46 2E Output peak voltage reading failed lower than expected Err 47 2F Output peak voltage reading failed higher than expected Err 48 30 Output peak current reading failed lower...

Page 26: ...ins Fuses T 5 A Monopolar mode with the possibility to connect 1 or 2 electrodes holder handles both of them with hand switches or just one with hand switches and the other one with pedal foot switche...

Page 27: ...Directive 89 336 CEE Category A Suggested distances to keep from not vital devices Source of the Current RF Typical Power W Distance m Microcellular telephones CT1 CT2 CT3 0 01 0 4 Mobile telephones D...

Page 28: ...300NT_EN B 18 04 2018 ssta cgro cgro 28 von 35 MONOPOLAR CURRENTS Variation of the output power from 50 to 2000 by selecting 100 and 50 of maximum power Increase in the output power with nominal load...

Page 29: ...50 200 250 300 3500 3000 2500 2000 1000 500 1500 20 40 60 80 100 120 140 7000 6000 5000 4000 3000 2000 1000 7500 Vpp V 50 100 50 100 50 100 P W RMS Impostazione Reglage P W RMS Setting Impostazione Re...

Page 30: ...120 150 5000 4000 3000 2000 1000 50 100 150 200 250 3500 3000 2500 2000 1500 1000 500 50 100 50 100 50 100 200 300 400 500 600 700 800 900 1k 1 1k 1 2k 1 3k 1 4k 1 5k 1 6k 1 7k 1 8k 1 9k 2k 100 200 3...

Page 31: ...0 200 240 280 3500 3000 2500 2000 1500 1000 500 20 40 60 80 100 120 140 7000 6000 5000 4000 3000 2000 1000 7500 50 100 50 100 200 300 400 500 600 700 800 900 1k 1 1k 1 2k 1 3k 1 4k 1 5k 1 6k 1 7k 1 8k...

Page 32: ...X 300NT_EN B 18 04 2018 ssta cgro cgro 32 von 35 BIPOLAR CURRENTS Variation of the output power from 10 to 1000 by selecting 100 and 50 of maximum power Increase in the output power with nominal load...

Page 33: ...40 20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 20 40 60 80 100 120 140 160 20 40 60 80 100 120 130 20 40 60 80 100 120 130 40 80 120 160 20 60 100 140 100 200 300 400 500 600 700 800 850 20 40...

Page 34: ...20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 20 40 60 80 100 120 130 20 40 60 80 100 120 130 20 40 60 80 100 120 130 450 400 350 300 250 200 150 100 50 20 40 60 80 100 12...

Page 35: ...RIES see next page for details Nominal voltage 4000 V E 1 Straight knife electrode E 5 Straight needle electrode thick type 2 pcs E 7 Straight needle electrode thin type E12 Straight ball electrode 2...

Reviews: