The pre-setting of these memories allows to use the unit as follows:
A)
With 2 active electrodes:
One active electrode or by hand-switches
(connected to the socket
3-A
) or by foot-switches
( connected to the
socket
5-A
)
.
In order to activate the different active electrodes shall be
pressed (on the handles or on the twin foot switch)
the
yellow
key/pedal for the pure/coagulating cut and the
blue
key/pedal for the coagulation currents.
In order to use two electrodes-holder handles (as above described) at the same time by two different
operators,
it must be selected a coagulation mode without contact (
Fulg
or
Spray
), as stated by the
International Safety Rules for the HF electro-surgical equipments (IEC 60601-2-2, par. 46.103).
Another active electrode, for the normal electro-surgery by hand-switches (connected to the socket
4-B
)
pressing the
yellow
key for the pure/coagulating cut and the
blue
key for the coagulation currents (See Par.
“CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”);
C. !!! BIPOLAR FUNCTIONING !!!
The bipolar functioning is always available. See Par. “BIPOLAR MODE”.
PRE-SET MEMORIES (nr.4)
– Use in cardio surgery.
The pre-setting concerns:
1.
Activation modes and outputs available.
2.
Starting settings of the output powers for cardio-surgery (70W for all the monopolar cutting currents
PURE,
BLEND 1, BLEND 2
,70W for the monopolar coagulating currents
FORCED, PIN POINT, SPRAY setting
itself on SPRAY because it is the more appreciated by the cardio surgeons
)
3.
Activation modes and outputs available for cardio-surgery.
4.
Starting settings when the Argon section is switched on (Automatic selection of the SPRAY coagulation current
that is needed for the Argon coagulation “the Led in the area
D
blinks when this type of current is not the one
used also for the electrosurgical mode”, automatic delivery of the SPRAY current at 60W when the
blue
key
on the electrodes-holder handle for Argon gas (
34-9730
) is pressed, automatic setting and delivery of the Argon
gas flow at the level of “5-6 lt/min”).
The activation modes and the outputs available can not be changed, but all the currents, the output powers
and the Argon gas flow (if it is used) can obviously be changed by setting the desired levels and by pressing
the memorisation key MEM of the area B.
The pre-setting of these memories allows to use the unit as follows:
A)
With 2 active electrodes:
One active electrode or by hand-switches
(connected to the socket
3-A
) or by foot-switches
( connected to the
socket
5-A
)
.
In order to activate the different active electrodes shall be
pressed (on the handles or on the twin foot switch)
the
yellow
key/pedal for the pure/coagulating cut and the
blue
key/pedal for the coagulation currents.
In order to use two electrodes-holder handles (as above described) at the same time by two different
operators,
it must be selected a coagulation mode without contact (
Fulg
or
Spray
), as stated by the
International Safety Rules for the HF electro-surgical equipments (IEC 60601-2-2, par. 46.103).
Another active electrode, for the normal electro-surgery by hand-switches (connected to the socket
4-B
)
pressing the
yellow
key for the pure/coagulating cut and the
blue
key for the coagulation currents.(See Par.
“CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES”);
C. !!! BIPOLAR FUNCTIONING !!!
The bipolar functioning is always available. See Par. “BIPOLAR MODE”.
PRE-SET MEMORIES (nr. 5)
– Use in open sky/laparoscopic surgery.
The pre-setting concerns:
1.
Activation modes and outputs available.
2.
Starting settings of the output powers (80W for all the monopolar cutting currents and 70W for all the monopolar
coagulating currents)
3.
Activation modes and outputs available.
The activation modes and the outputs available can not be changed, but all the currents and the output
powers can obviously be changed by setting the desired levels and by pressing the memorisation key MEM
of the area B.
The pre-setting of these memories allows to use the unit as follows:
A)
With 2 active electrodes:
One active electrode or by hand-switches
(connected to the socket
3-A
) or by foot-switches
( connected to the
socket
5-A
)
.
In order to activate the different active electrodes shall be
pressed (on the handles or on the twin foot switch)
the
yellow
key/pedal for the pure/coagulating cut and the
blue
key/pedal for the coagulation currents.
In order to use two electrodes-holder handles (as above described) at the same time by two different
operators,
it must be selected a coagulation mode without contact (
Fulg
or
Spray
), as stated by the
International Safety Rules for the HF electro-surgical equipments (IEC 60601-2-2, par. 46.103).