This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key
MODE
, select the current
BLEND-1
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the twin foot-switch
pedal
Instructions for use, adjustments, and electrodes
As it is detailed for the current PURE.
COAGULATING CUT (BLEND-2)
– Not suitable at all.
COMBINATED CUT (CUT ALTERNATED TO COAGULATION)
(ENDO)
– Not suitable at all.
CURRENTS FOR COAGULATION
“FULGURATION” COAGULATION AT HIGH VOLTAGE - Strong superficial sparkling effect, and
optimum deep coagulating effect
This current guarantees a strong coagulating effect, both deep and superficial. It is suitable
both for coagulations performed with forceps/surgical instrument, and for coagulations
performed by grazing the tissues with the active electrode.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key, select the current
FULG
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
acute
acoustic signal, and
blue
Led), press on the
blue
pedal of the twin foot-switch.
Instructions for use, adjustments, and electrodes
Use the loop/ball electrodes for the resectoscope from 60-70W.
“PIN POINT” COAGULATION AT MEDIUM VOLTAGE – Medium-Low superficial sparkling effect, and
optimum deep coagulating effect
This current guarantees a strong deep coagulating effect and a normal superficial coagulating
effect. It is suitable both for coagulations performed with forceps/surgical instrument, and for
coagulations performed by grazing the tissues with the active electrode, just in case the
operators prefer a superficial effect which is less strong than the one of the FULG coagulation.
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key
MODE
, select the current
PIN POINT
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
acute
acoustic signal, and
blue
Led), press on the
blue
pedal of the twin foot-switch .
Instructions for use, adjustments, and electrodes
Use the loop/ball electrodes for the resectoscope from 60-70W.
“SOFT” COAGULATION AT LOW VOLTAGE
– Not suitable.
“SPRAY” COAGULATION AT VERY HIGH COLTAGE (
SPRAY
)
– Very strong superficial sparkling
effect, and good deep coagulating effect
This current guarantees a good deep coagulating effect, and a very strong superficial
coagulating effect. It is suitable both for the coagulations performed with forceps/surgical
instrument, and for coagulations performed directly with the active electrode, even far away
from the tissues.