Alternatively, two active electrodes
(two handles, or one handle and one forceps for coagulation, or other),
both of them by foot-switches
(This mode is possible by selecting the pre-
set programs 6 and 7 “See Par.
“PROGRAMS AND MEMORIES”);
E.
!!! BIPOLAR FUNCTIONING !!!
The unit can always be used for the bipolar functioning. See Par. “BIPOLAR MODE”.
In order to favour this kind of use, the unit is equipped with 3 specific pre-set programs (See Par.
“PROGRAMS AND MEMORIES”, programs from 3 to 5), where the following features are already
memorized:
1.
Activation modes and outputs available
2.
Starting settings of the output powers (1 program for cardiosurgery, 1 program for neurosurgery, 1 program for
general/different surgeries)
In all cases, it is possible to change all the above specified pre-settings, according to the personal
need/desire, and therefore memorize these variations into the different programs (See Par.
“PROGRAMS
AND MEMORIES”).
MONOPOLAR CURRENTS, ELECTRODES, ADJUSTMENT OF THE POWERS, ADVICE
CURRENTS FOR CUT AND COAGULATING CUT
PURE CUT (PURE)
– Cut without coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key
MODE
, select the current
PURE
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the twin foot-switch, or
the
yellow
key of the hand-switch handle.
Instructions for use, adjustments and electrodes
Use cutting electrodes (blade, needle, loop, Hook type, or the external part of the dissector, in laparoscopy) from
30-40W.
COAGULATING CUT (BLEND-1)
– Cut with soft coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key, select the current
BLEND-1
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the foot-switch, or the
yellow
key of the hand-switch handle.
Instructions for use, adjustments and electrodes
Use cutting electrodes (blade, needle, loop, Hook type, or the external part of the dissector, in laparoscopy) from
30-40W.
COAGULATING CUT
(BLEND-2)
– Cut with a very strong spray coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key, select the current
BLEND-2
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the foot-switch, or the
yellow
key of the hand-switch handle.
Instructions for use, adjustments and electrodes