FUNCTIONING FOR THE FLEXIBLE ENDOSCOPIC SURGERY
In order to favour this kind of use, the unit is equipped with 1 specific pre-set program (See Par.
“PROGRAMS AND MEMORIES”, program 8), where the following features are already memorized:
1.
Activation mode and output available
2.
In this program are not set output powers or currents for the normal monopolar use because this choice
depends on the operator.
3.
In all cases, it is possible to set standard pre-set currents/output powers and therefore memorize these
settings into the program (See Par.
“PROGRAMS AND MEMORIES”).
CONNECTION AND USE OF THE ACCESSORIES
1. Socket (5-A, the led A under is lighted on): flexible electrode for electro-
surgery with pedal foot-switch
(See Par. “CONNECTION AND USE OF
THE PEDAL FOOT-
SWITCHES”).
Use only this socket to connect the above mentioned instruments. The use of different sockets provokes a
damage to the unit.
Just in case, ask for:
The adaptors for the cables with different plug:
Adaptor
-
34-9590
, for cables with non insulated plugs
from 2 to 8mm, or with insulated plug
4mm;
2. Socket (7- NEUTRAL PLATE): neutral electrode.
According to the connection possibilities listed above, the unit allows to:
A. Use the flexible electrode for the normal HF electro-surgery by foot-switch activation
(See Par. “
CONNECTION AND USE OF THE PEDAL FOOT-SWITCHES
” ). Press on the
yellow
pedal for the
pure/coagulating
cut currents, and the
blue
pedal for the coagulation currents.
C. !!! BIPOLAR FUNCTIONING!!!
The bipolar functioning is always possible. See Par.
“BIPOLAR MODE”.
MONOPOLAR CURRENTS, ELECTRODES, ADJUSTMENT OF THE POWERS, ADVICE
CURRENTS FOR CUT
PURE CUT
– Cut without coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key
MODE
, select the current
PURE
, then adjust the power with the regulation keys
UP/DOWN
;
- For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal
of the twin
foot-switch
Instructions for use, adjustments and electrodes
Use the flexible electrodes of the endoscopes, from 20-30W for the papillotomy, from 20-30W for the
polipectomy with polypus
5mm, and from 40-50W for polypus
6mm or more.
COAGULATING CUT (BLEND-1)
– Cut with soft coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area C):
-
With the selection key, select the current
BLEND-1
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal
of the twin foot-switch.
Instructions for use, adjustments and electrodes