This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area D):
-
With the selection key
MODE
, select the current
SPRAY
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
acute
acoustic signal, and
blue
Led), press on the
blue
pedal of the twin foot-switch.
Instructions for use, adjustments, and electrodes
Use the loop/ball electrodes for the resectoscope from 60-70W.
ADVICE
For the Cut with wire loop electrodes, use as follows:
Current
PURE,
from 70-80W into the bladder or on soft tissues, for a cut without coagulating effect;
Current
BLEND-1
, from 70-80W into the bladder or on soft tissues, for a cut with soft coagulating effect;
Current
PURE,
from 90-100W on the prostate, for a cut without coagulating effect;
Current
BLEND-1
, from 90-100W on the prostate, for a cut with soft coagulating effect.
For the Cut with ribbon loop electrodes, use as follows:
Current
PURE,
from 160-180W into the bladder or on soft tissues, for a coagulating cut;
Current
BLEND-1
, from 160-180W into the bladder or on soft tissues, for a cut with stronger coagulating effect;
Current
PURE,
from 200-210W on the prostate, for a coagulating cut;
Current
BLEND-1
, from 200-210W on the prostate, for a cut with stronger coagulating effect.
NB: In order to optimize the functioning and obtain the perfect cut without any “sticking” of the tissues on
the electrode, the above mentioned powers maybe have to be changed through small variations of 10-15W
each time.
For the Vaporisation with rolls, use as follows:
Current
PURE,
from 100-120W with small rolls, and from 200-210W with big rolls.
For the Coagulation, use as follows:
Current
SPRAY
from 40-50W. Optimum to coagulate or by loop/blade/ball electrodes, even without any contact
with the tissues;
Current
FULG,
from 60-70W, for a coagulation by loop/blade/ball electrodes.
Current
PIN POINT
, from 60-70W, for a coagulation by loop/blade/ball electrodes with contact between the
electrode and the tissues.