BIPOLAR MODE
INTRODUCTION
As far as the bipolar functions, please note as follows:
-
They are always available, as alternative to all the other monopolar functioning modes;
-
They can be used alone, without connecting the neutral electrode. See. the pre-set programs in next paragraph;
-
Besides the standard twin foot-switch (
93-9330
, for the monopolar or bipolar use), the unit can also be equipped
or with another twin foot-switch (
34-9331,
for the bipolar use only), on request. (See Par. “CONNECTION AND
USE OF THE PEDAL FOOT-
SWITCHES”);
-
The unit is equipped with a coagulating current
AUTO
(impedance sensing type) with automatic start/stop
system (switching on with adjustable delay from 0.5 to 5sec by the control device
b5
which is placed on the
backside of the unit, when the tips of the forceps get in touch with the non coagulated tissues
– switching off
when the tissues are coagulated). This system does not require any pedal foot-switch;
CONNECTION AND USE OF THE ACCESSORIES
The unit is equipped with one socket for the connection of the bipolar electrodes (
8-BIPOLAR)
.
1.
Socket 8
- BIPOLAR-A
: bipolar electrode (connected cable without any particular polarity) for use
by foot-
switch
for cutting and/or coagulation. The foot-switch for the
AUTO
coagulation with automatic start/stop
system is not used;
In order to facilitate the bipolar use, the unit is equipped with 1 pre-set program
(See Par. “PROGRAMS AND
MEMORIES”,
program 9) where the following features are already memorized
:
-
Disconnection of the neutral electrode safety circuit, which is not required for the bipolar use only (as a matter
of fact, through the 3 lit alarm Led, the circuit just informs the operators that the neutral electrode is not
connected);
-
Impossibility to use the monopolar currents (the displays for the monopolar currents powers are switched off);
-
Activation modes and outputs available according to the different settings
Just in case of cables with connection plugs which are different from GEISTER standard type, ask for:
-
GEISTER cables
, by specifying the model and the instrument connector type;
-
Adaptor
,
34-9591V
for cables with connector (Valleylab type).
-
Adaptor
,
34-9591E
for cables with connector (coaxial type, Erbe or Storz type).
-
Adaptor
,
34-9591M
for cables with connector (coaxial type, Martin or Bertchold type).
BIPOLAR CURRENTS, ELECTRODES, ADJUSTMENT OF THE POWERS, ADVICE
CURRENTS FOR CUT
PURE CUT
– Cut with minimum coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area L):
-
With the selection key
MODE
, select the current
PURE
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the twin foot-switch.
Instructions for use, adjustments, and electrodes
Use the specific electrodes, from:
40-50W, with cutting electrodes, needle type;
COAGULATING CUT
– Cut with coagulating effect
This current is controlled by an automatic system (
ADC System
) for the self regulation of the
functioning, which guarantees that the power selected by the operators remains constant
according to the characteristics of the tissues.
Control devices to be used (area L):
-
With the selection key
MODE
, select the current
BLEND
, then adjust the power with the regulation keys
UP/DOWN
;
-
For the delivery (
loud
acoustic signal, and
yellow
Led), press on the
yellow
pedal of the twin foot-switch.
Instructions for use, adjustments, and electrodes
As it is detailed for current PURE.