GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm
Page no. 181
Chapter 16
CC8
Reallocation of disk space
failed. There is no space
available on disk for
acquiring more images.
Disk system is full or server has
failed.
1. Close exam.
2. Increase free space on disk system by
deleting exams that have been archived.
3. Perform system shutdown, then
restart system.
4. If the problem persists, contact
your
Field Service Representative
.
CC9
System disk is full
System disk space is full.
1. Close exam.
2. Increase free space on disk system by
deleting exams that have been archived.
3. Perform system shutdown, then
restart system.
4. If the problem persists, contact
your
Field Service Representative
.
CC10 The DICOM image con-
struction failed.
The system is not able to format
acquired image.
1. Close exam.
2. Increase free space on disk system by
deleting exams that have been archived.
3. If the problem persists, contact
your
Field Service Representative
.
Message on AWS screen
Explanation