GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm
Page no. 183
Chapter 16
2D Large Localization
Paddle. Maximum rec-
ommended force is
20daN.
The system informs the operator
that the "Biopsy" paddle is present
and reminds the operator that it is
recommended to not exceed
20daN with this paddle.
If the "Biopsy" paddle is not installed, an error has
occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
No defined paddle
The system informs the operator
that it does not detect a
Senographe DS paddle.
If you observe this message when a Senographe
DS paddle is present, contact GEMS Field Service.
Round Spot Magnifica-
tion paddle
The system informs the operator
that the "Round Mag Stand" pad-
dle is present. This paddle should
be used with a magnification plat-
form.
If the "Round Mag Stand" paddle is not installed,
an error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Axillary paddle
The system informs the operator
that the "Axillary" paddle is pres-
ent.
If the "Axillary" paddle is not installed, an error has
occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
19x23 paddle
The system informs the operator
that the "19x23 Standard" paddle
is present.
If the "19x23 Standard" paddle is not installed, an
error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Flexible 19x23 paddle
The system informs the operator
that the "19x23 Phlexis" paddle is
present.
If the "19x23 Phlexis" paddle is not installed, an
error has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Round spot paddle
The system informs the operator
that the "Round spot" paddle is
present.
If the "Round spot" paddle is not installed, an error
has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
2D Spot Localization
Paddle. Maximum rec-
ommended force is
20daN
The system informs the operator
that the "Spot Biopsy" paddle is
present and reminds the operator
that it is recommended to not
exceed 20daN with this paddle.
If the "Spot Biopsy" paddle is not installed, an error
has occurred:
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages
and contact GEMS Field Service.
Square Magnification
paddle
The system has detected that the
2D Spot localization paddle is
present and reminds the Operator
that it is recommended to not
exceed 20daN with this paddle.
If the 2D Spot localization paddle is not installed,
an error has occurred:
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages
and contact your Field Service Representative.
Message on
Gantry Readout
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).