Chapter 4
Page no. 38
4-cons.fm
GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
X-ray Console
3-1
Access to setup menus
DEUTS
ENGL
FRANC
PORTU
ITALI
ESPAN
•
With the X-ray Console in normal application mode, press the
Setup
button to enter the "menu tree"
shown below.
•
To access a function, press the button below the displayed name of the function.
•
To back out of a function (move "up the tree"), press the
Setup
button again.
view medical language PASSWORD
DECOMP COMP LIFT BEEP ROTATE STEREO
+ - VALID
FAST SLOW
YES NO
recumbent
medical set up
DECOMPRESSION HEIGHT: 21 cm
END OF EXPOSURE DECOMP:
DECOMP HEIGHT
LIGHT
MAX COMPRESSION FORCE: 16 daN
FORCE SPEED
PRESET SPEED
FAST SLOW
FAST SLOW
COMPRESSION SPEED: FAST
DEFAULT OBLIQUE ANGLE: 45 deg
DEFAULT LIGHT DURATION: 3 min
LIFT FOOTSWITCH SPEED: SLOW
WITH BEEP: YES
ROTATION SPEED: SLOW
+ - VALID
+ - VALID
+ - VALID
YES NO
+ - VALID
Function available for service only
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).