Chapter 16
Page no. 174
16-errors.fm
GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
S25 Tube housing over
temperature
The temperature of the tube
housing is too high (> 65°C).
Wait for the tube housing temperature to fall.
S26 Exposure too long
The exposure is too long (this
message should appear only in
Manual mode).
Increase the kV value or reduce the mAs value.
S32 Compression sen-
sor fault
Breast thickness is badly calcu-
lated in AOP mode.
It is still possible to continue the
exam in Manual mode.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
S33 Wait Until End of
Digital Acquisit.
Acquisition is inhibited until the
previous image has been com-
pletely recorded.
No action is required. The display of this message
is normal during image transfer to the workstation.
It will be cleared when system is ready for the next
acquisition.
If the problem persists, power OFF/ON from the X-
ray Console.
S37 Exp. Abort, Wait
Image Acquisition
Exposure in progress has been
aborted by the operator or the
equipment but the system pro-
cessed and displayed the gener-
ated image.
1. See the message displayed when the Exposure
Interrupt button is lit for more information.
2. Press the Exposure Interrupt button and perform
the action required by the last displayed message.
S38 Abort by HV arcing,
Wait Image Acq.
Exposure was aborted because
more than eight X-ray tube arcing
events occurred, but the system
processed and displayed the
generated image.
1. See the message displayed when the Exposure
Interrupt button is lit for more information.
2. Press the Exposure Interrupt button and perform
the action required by the last displayed message.
S39 Abort by IDC, Wait
Image Acquisition
The exposure was aborted by the
Image Detection Controller, but
the system processed and dis-
played the generated image.
1. See the message displayed when the Exposure
Interrupt button is lit for more information.
2. Press the Exposure Interrupt button and perform
the action required by the last displayed message.
S40 Stereo Config/
Exam incoherence
After power reset on the X-ray
Console, the system has not
retrieved the correct configura-
tion.
Request a new stereotactic series by clicking the
"New Series" button on the Viewer. The scout
image must be acquired again.
S50 SP present, cannot
continue exam
Stereotactic Positioner was
detected as present with the
Routine Medical Application
open.
1) If Stereotactic Positioner is present:
a) To open the Stereo Medical Application, first
close the current exam.
b) To use the Routine Medical Application, first
remove the Stereotactic Positioner.
2) If Stereotactic Positioner is not present:
a) Power OFF/ON from the X-ray Console.
b) If the problem persists, shut down the system
from the Browser.
c) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
X-ray Console
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!