GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
152
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
I56
GAN
Slipping
Compression
Motor Slipping
Compression motor
has slipped and lost its
position.
1. Release footpedal or compression
button, and move paddle up to park
position.
2. Repeat the compression.
I62
GAN
Oscillations
Arm oscillations
Gantry oscillate then
the sensor angle value
is not stable.
This angle variation
could affect view name
calculation.
1. Wait that the gantry stabilizes.
I70
GAN
Comp
Failure
Compression
Failure
Compression did not
function correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I71
GAN
Light Fail
Light Centering
Device Fail
The light centering
device is not
functioning correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I72
GAN
Column Fail
Column Motion
Failure
Elevator is not
functioning correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I73
GAN
Rotat. Fail
Rotation Brake
Failure
Rotation brake is not
functioning correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I74
GAN
Lock Failure
Arm Lock Failure
Arm lock is not
functioning correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I75
GAN
Button Fail
Button Security
Failure
Problem reported in
operator presence
interlock in control
buttons.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I76
GAN
Failure
Arm Sensor
Failure
An error has occurred
in the rotation angle
sensor.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.
I77
GAN
Pedal Fail
Comp Pedal
Security Failure
The operator presence
interlock for the
compression pedal did
not function correctly.
Switch OFF Gantry, then switch ON.
If problem continues, note ALL
messages and contact GEMS Field
Service.