Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
S88
AWS
None
S88 AWS image
delivery failure
AWS image delivery to
medical application
default.
1. If this message stays displayed
more that 5mn switch OFF the whole
system and wait for the end of the
shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S89
AWS
None
S89 AWS not
available for
acquisition
The AWS cannot allow
acquisition because of
lack of disk space or
an uncomplete action
on AWS.
1. Check that all action are complete
2. Increase free space on image disk
by deleting exams that have been
archived.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S90
AWS
None
S90 AWS not
available, close
exam
No more AWS
ressources to continue
examlem occurs in
AWS.
1. Close exam.
2. Increase free space on image disk
by deleting exams that have been
archived.
3. If the problem persists, note ALL
messages and inform GEMS Field
Service.
S91
AWS
None
S91 Need user
action on AWS
Operator input is
required on AWS.
Answer to AWS pop–up.
S92
AWS
None
S92 AWS busy,
please wait...
The AWS is busy, and
the current operation
may take some time.
No action required, wait for
completion of the current operation.
S93
AWS
None
S93 AWS busy
The AWS is busy (the
wait should be short).
No action required, wait for
completion of the current operation.
S94
AWS
None
S94 System
performing test,
wait
An application is
running on the system
and performs tests that
should not allow
acquisition (example
auto–diagnostics).
Wait few minutes.