CHAP
. 14
GE Healthcare
Senographe 2000 D Acquisition System
REV 1
OM 5179217–1–100
156
Msg.
No.
Corrective Action
Explanation
Message on
Control Console
Message
on Column
Readout
Source
S26
GEN
None
Exposure too
long
The exposure is too
long (this message
should appear only in
manual mode).
Increase the kV value or reduce the
mAs value.
S32
GEN
None
Compression
sensor fault
Breast thickness is
badly calculated in
AOP mode.
It is still possible to
continue the exam in
manual mode.
Switch OFF Gantry, then switch ON.
If the problem continues, note ALL
messages and contact GEMS Field
Service.
S33
GEN
None
Wait Until End of
Digital Acquisit.
Acquisition is inhibited
until the previous
image has been
completely recorded.
No action is required. The display of
this message is normal during image
transfer to the workstation. It will be
cleared when system is ready for the
next acquisition.
If the problem persists, switch OFF
gantry, then switch it ON.
S37
GEN
None
Exp. Abort, Wait
Image Acquisition
Exposure in progress
has been aborted by
the operator or the
equipement but the
system processed and
display the image that
has been generate.
1. Look at the message displayed
when the reset button is highlighted
to know the real problem
2. Press reset and perform the action
required by the last displayed
message.
S38
GEN
None
Abort by HV
arcing. Wait
Image Acq.
S39
GEN
None
Abort by IDC.
Wait Image
Acquisition
S50
GAN
Expos
Impos
Incorrect X_Ray
Incidence
The position of the
tube carrying arm does
not allow exposure: –
Position not locked at
0 degrees.
Correct the position of the tube
carrying arm.
S81
AWS
None
S81 Image
detection not
ready
One of the image
detection coponent is
buzy or its condition to
acquire images is not
yet reached.
1. Wait for completion of system
setting.
2. If this X–ray inhibition stays
displayed more that 10 mn Switch
OFF the whole system and wait for
the end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL
messages and inform GEMS Field
Service.