CHAPTER 6 – START UP AND SHUTDOWN
1.
START UP PROCEDURE
To power up the system, press the power switch
I
on the Control Console. Power is applied to the
whole system (Gantry, Cart, Digital Detector, etc.); green power indicators are lit on the Generator, the
Cart, and the monitor.
A series of system initialization messages appears on the Workstation screen, then the Login: prompt
is displayed. Type
sdc
(all lower case) followed by
<Enter>
. When the password prompt appears, type
adw3.1<Enter>
. The AWS Browser is displayed after a short delay.
2.
SYSTEM SHUTDOWN
To power down the system, click the System Shutdown button
on the AWS Browser window. A
confirmation message is displayed asking if the entire system should be shut down, If the response is
Yes
, power is removed from the Senographe 2000 D Gantry and other system components, while the
Acquisition Workstation begins a controlled shutdown, designed to safeguard data. Complete system
and UPS shut down takes about two minutes from switch-off.
The Acquisition Workstation follows an automatic shut down procedure which protects
image and patient data. No prompts are displayed. Do not disturb this procedure by
attempting to enter commands at this time.
NEVER switch off at the UPS (Uninterruptible Power Supply) except in emergency
(risk of data loss).
3.
EMERGENCY SHUTDOWN
D
In an emergency, to cut power to the Gantry:
Press one of the Emergency Stop push-buttons;
there is one on each side of the Telescopic Column.
D
In an emergency, to cut power to all system components
: Switch off the main circuit breaker
that supplies power from the hospital system to the Senographe 2000 D system. All image and
patient information being processed at the time is lost, and the Conditioner which controls the
Digital Detector environment is switched off.
4.
GANTRY RESET
Certain configuration operations require the Gantry to be switched off and restarted. It is usually
desirable to reset the Gantry without disturbing the operation of the Acquisition Workstation.
To do this, switch off the Gantry and Generator by pressing the Control Console power switch
STOP
or
.
Switch them on again by pressing the power switch
I
.
5.
GENERATOR CABINET PUSH-BUTTONS
Push-buttons mounted on the Generator cabinet control power to the
AWS Cart
and to the
UPS
. They
are provided for service and emergency use; they should not be used in normal operation.
NOTICE
CAUTION
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).