Programming with PanaView
4-11
March 2008
Entering User Tables
To support user functions, the
MMS-3
can hold up to five tables
(designated as A through E) of non-linear or empirical data. Users can
enter up to 16 X-Y pairs in each table. A user function can supply an
X value with Tbl(). The
MMS-3
then interpolates the Y value for a
given X, and substitutes it for Tbl (X) in the function.
To enter values for a user table:
1.
From the “
New Meter Browser”
option, click on the desired
communications port and meter.
2.
Click on “
User Tables
,” and then double-click on the desired
table.The right side of the screen appears similar to Figure 4-8
below.
Figure 4-8: Blank User Table
3.
Enter a name (up to 13 characters) for the table in the “Table
Name” text box, and enter up to 16 X and Y values in the
appropriate columns.The table now appears similar to Figure 4-9
below.
Figure 4-9: Completed User Table
4.
When you have finished entering values, click on
[Apply]
to apply
the table immediately, and on
[OK]
to confirm the table and close
the window. To return the window to its previous settings, click on
the
[Cancel]
button.