March 2008
2-4
Basic Programming
Displaying
Measurements (cont.)
When selecting a measurement mode, please remember that some
measurement modes require multiple inputs. For example, to display
ppm
v
, you need a moisture and pressure input. Table 2-2 below shows
a list of measurement modes that require multiple inputs and what
you need to display them. If the
MMS-3
displays “No Data,” you may
have to reconfigure the channel. Refer to
Reconfiguring a Channel
for a New Sensor
on page 3-35.
.
Repeat this procedure to set up the other line. When you are done,
press
[RUN]
.
Table 2-1: Measurement Modes
Display Abbreviation
Measurement Mode
O
Oxygen
H
Hygrometry
T
Temperature
P
Pressure
AUX1
Auxiliary 1
AUX2
Auxiliary 2
USER*
User Function
VREF
Volt Reference
GND
Signal Ground
*User functions are available only through PanaView
software. For more information, refer to Chapter 4.
Table 2-2: Measurement Modes and Required Inputs
To measure:
You need the following inputs:
RH
Temperature and moisture
PPM
v
Moisture and pressure
PPM
w
Moisture, temperature and
saturation constant data
PPB
v
Moisture and pressure
MCF/IG
Moisture and pressure
MCF/NG
Moisture and pressure
PPM
v
/NG
Moisture and pressure
Select HYGRO Unit L1
Use the arrow keys to move the
brackets to the measurement
units and press
[YES]
.
[DP/°C] DP/ °F
`