Advanced Programming
3-1
March 2008
Introduction
Now that the
MMS-3
is running and the screen is set up, you may
want to use the other operational features included in this chapter (see
the previous page for the list and locations of those features).
Setting Up the Alarms
The
MMS-3
can be equipped with an optional high and low alarm
relay for each channel.
Press the
[PROG]
key to enter the user program.
Note:
If you have already entered the user program, see Figure 3-3
on page 3-49 to navigate to the
Alarms
Menu.
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
[CHAN]
key to select the desired channel.
The
MMS-3
briefly flashes a message, “Relay Status -- Reset,” and
then displays the next prompt.
Enter Passcode: XXXX
Enter the passcode.
Programming Menu 1
Use the arrow keys to move the
brackets to
ALARMS
and press
[YES].
_
[ALARMS]
`
Select Alarm 1
Use the arrow keys to move the
brackets to the desired alarm
(
A
or
B
) and press
[YES]
.
[A] B
Select Mode 1
Use the arrow keys to move to
the desired measurement mode
and press
[YES]
. See Table 3-1 on
page 3-2 for a list of available
measurement modes.