Advanced Programming
3-39
March 2008
Entering Calibration
Data for New Probes/
Sensors
The
MMS-3
needs calibration data for each moisture, oxygen and
pressure sensor, as well as each auxiliary input. GE Sensing supplies
each probe with a Calibration Data Sheet, which is a list of 1 to 16
data points the electronics needs to interpret raw data from the probe.
Calibration Data Sheets are packed with each sensor.
Note:
If you want to enter a constant value for moisture,
temperature, pressure, or multiply the ppm
v
value by a
constant multiplier, refer to
Entering System Constants
on
page 3-26. It is not necessary to enter the calibration data if
you are using a constant.
Note:
If you have already entered the System Menu, and selected a
channel and measurement mode, proceed to the CURVES
selection step.
Press the
[PROG]
key to enter the user program..
Be sure the number displayed in the upper right-hand corner of the
screen is the channel you want to program. If not, press the
[CHAN]
key to select the desired channel.
.
Enter Passcode: XXXX
Enter the passcode
.
Programming Menu 1
Use the arrow keys to move the
brackets to
SYSTEM
and press
[YES].
[SYSTEM] AUTOCAL
`
Measurement Mode 1
You must now select the type of
measurement the probe is
capable of taking. Use the arrow
keys to move to the desired
measurement mode and press
[YES].
page 3-40 for a list of available
measurement modes.
O [H] T P AUX1
`