March 2008
3-14
Advanced Programming
Checking the Battery
Status
If your
MMS-3
has the battery pack, you should periodically monitor
the amount of charge the battery has left, as described below.
Disconnect the AC power. Press the
[PROG]
key to enter the user
program.
Note:
If you have already entered the user program, see Figure 3-3
on page 3-49 to navigate to the Test Menu.
Enter Passcode: XXXX
Enter the passcode
.
Programming Menu 1
Use the arrow keys to move the
brackets to
TEST
and press
[YES]
.
_
[TEST] CONTRAST
`
Test Menu 1
Use the arrow keys to move the
brackets to
BATTERY
and press
[YES]
.
_
[BATTERY]
Battery Test
Use the arrow keys to move the
brackets to
STATUS
and press
[YES]
.
[STATUS] RdChgTime
Battery Status (v)
The
MMS-3
alternates two
displays showing the battery
voltage and how much longer the
battery will operate (in minutes).
XX.XX Press [YES]
Run Time Left (min)
XXX