Entering Pressure Curve
Data
Repeat the three previous steps to enter the remaining data.You must
enter the necessary data for zero and span.
You may now do one of the following:
•
Enter data for a different channel or type of sensor by pressing the
[NO]
key until you reach the Programming Menu, and then refer to
Entering Calibration Data for New Probes/Sensors
on page 3-39.
•
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
•
Exit by pressing the
[RUN]
key.
IMPORTANT:
You must configure a channel AND enter the
necessary calibration data in order for a sensor to
work properly.
Pressure Curve Menu 1
Use the arrow keys to move the
brackets to
S/N
and press
[YES]
.
[S/N] CURVE
Probe Serial # 1
Enter the serial number and press
[YES].
Then press
[NO]
to return to
the Curve Menu.
P: 123456
Pressure Curve Menu 1
Use the arrow keys to move the
brackets to
CURVE
and press
[YES].
S/N [CURVE]
# mA Psig 1
Making sure the cursor is
positioned below the # sign,
enter a 1 (for zero) or 2 (for
span). Press
[YES]
and then press
the right arrow key.
1 +0.000 0.000
# mA Psig 1
Enter the mA value. Press
[YES]
and press the right arrow key.
1 +1.000 0.000
# mA Psig 1
Enter the psig value. Press
[YES].
1 +0.000 2.000