Advanced Programming
3-33
March 2008
Setting Error Processing
(cont.)
If you selected one of the
ALARMS
:
Refer back to the “Selecting Error Response” step to continue error
processing.
Measurement Mode 1
You must now select the type of
measurement the probe is
capable of taking. Use the arrow
keys to move to the desired
measurement mode and press
[YES].
See Table 3-8 below for a
list of available measurement
modes.
O [H] T P Aux1
`
Table 3-8: Measurement Modes
Display Abbreviation
Measurement Mode
O
Oxygen
H
Hygrometry
T
Temperature
P
Pressure
AUX1
Auxiliary 1
AUX2
Auxiliary 2
CNST-PPMV
PPMv Multiplication Factor
System Menu 1
Use the arrow keys to move the
brackets to
ERROR
and press
[YES].
_
CONSTANT [ERROR]
Sel. ERROR Type 1
Use the arrow keys to move the
brackets to the desired selection
and press
[YES]
.
[LoRange] HiRange
Sel. ERR Response 1
Use the arrow keys to move the
brackets to the desired alarm or
recorder and press
[YES].
[ALM-A] ALM-B
`
Sel. ERR Alarm 1
Use the arrow keys to select
whether you wish to trip or
ignore the error and press
[YES].
[IGNORE] TRIP