Advanced Programming
3-29
March 2008
Entering a Saturation
Constant (cont.)
Repeat the last three steps to enter all the information for each data
point (maximum of six points). Enter the next data point number you
want to enter over the present data point number and press the right
arrow key. When you do this, the T/°C and Cs values will display
values of the current saturation constant curve. This indicates you are
ready to begin entering information for the selected data point.
You may now do one of the following:
•
To enter a saturation constant for the other channel, press
[NO]
until
you return to the Programming Menu on page 3-28.
•
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
•
Exit by pressing the
[NO]
key followed by the
[RUN]