March 2008
3-2
Advanced Programming
Setting Up the Alarms
(cont.)
Note:
The number of measurement choices and the units of
measurement vary depending on the measurement mode
selected.
Table 3-1: Measurement Modes
Display Abbreviation
Measurement Mode
O
Oxygen
H
Hygrometry
T
Temperature
P
Pressure
Aux 1
Auxiliary 1
Aux 2
Auxiliary 2
User*
User Function
Vref
Volt Reference
Gnd
Signal Ground
*User functions are available only through PanaView
software (see
Programming User Functions
in Chapter 4).
Select Unit 1
Use arrow keys to move the
brackets to the desired units and
[DP/°C] DP/ °F
`
press
[YES].
Select Trip Type 1
Use the arrow keys to move the
brackets to the desired trip type
(whether you want the alarm to
trip above or below the
programmed value) and press
[YES].
[Above] Below
A Setpoint DeadBD 1
Use the numeric keys to enter the
setpoint (the value above or
below which you want the alarm
to trip) and press
[YES]
to
confirm the entry. If you wish to
enter a deadband value
(explained on the next page),
press the right arrow key to reach
the value, use the numeric keys
to enter the value, and press
[YES].
+0.0 +0.0