Maintenance
Bottom Cover
The following procedures describe how to remove and replace the bottom cover.
Removing the Bottom Cover
Use the following procedure to remove the bottom cover.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the top cover assembly.
See
“Removing the Top Cover Assembly” on page 75
for more information.
3.
Remove the power supply assembly.
See step
through step
in
“Removing the Power Supply Assembly” on page
for instructions.
4.
If it is installed, remove the KISS pump and tubing.
See step
through step
in
“Removing the KISS Pump Assembly” on page 126
for instructions.
5.
Detach the chassis from the bottom cover.
See step
through step
in
“Detaching the Chassis” on page 82
for instructions.
6.
Remove the printer module assembly.
See step
through step
in
“Removing the Printer Module Assembly” on page
for instructions.
7.
Remove the printer door.
See step
through step
in
“Printer Door” on page 103
for instructions.
8.
Remove the power inlet module.
See step
through step
in
“Removing the Power Inlet Module” on page 93
for
instructions.
9.
Remove the four M3x5 fasteners securing the side and rear paper spacers to the
bottom cover, identified in the following photograph, and remove the spacers.
The bottom cover has now been successfully removed from the device and is
ready to be replaced.
2053535-003C
MAC™ 2000 ECG Analysis System
89
Summary of Contents for MAC 2000
Page 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Page 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Page 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Page 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Page 183: ......