Maintenance
Printer Door
The following procedures describe how to remove and replace the printer door.
Removing the Printer Door
Use the following procedure to remove the printer door.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the top cover assembly, if it is has not been removed already.
See
“Removing the Top Cover Assembly” on page 75
for more information.
3.
Detach the chassis.
See
“Detaching the Chassis” on page 82
for instructions.
4.
Remove the printer module assembly.
See
“Removing the Printer Module Assembly” on page 94
for instructions.
5.
Gently turn the device over so the bottom cover is facing you.
6.
Remove the four screws securing the printer door to the bottom cover, as shown
in the following photograph.
7.
Turn the bottom cover over and remove the dampers on the either side of the
printer door, as shown in the following photograph.