Maintenance
Removing the Power Inlet Module
Use the following procedure to remove the power inlet module.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the top cover assembly, if it is has not been removed already.
See
“Removing the Top Cover Assembly” on page 75
for more information.
3.
Detach the chassis.
See
“Detaching the Chassis” on page 82
for instructions.
4.
Remove the two M3x8 screws securing the power inlet module to the bottom
cover, as shown in the following photograph.
5.
Remove the power inlet module.
Replacing the Power Inlet Module
Use the following procedure to replace the power inlet module.
1.
Position the power inlet module in the bottom cover.
2.
Secure the power inlet module to the bottom cover using two M3x8 screws, as
shown in the following photograph.
3.
Reattach the chassis.
See step
through step
in
“Attaching the Chassis” on page 85
for instructions.
4.
Reattach the top cover assembly.
See
“Reattaching the Top Cover Assembly to the Device” on page 81
for
instructions.
5.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
6.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
2053535-003C
MAC™ 2000 ECG Analysis System
93
Summary of Contents for MAC 2000
Page 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Page 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Page 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Page 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Page 183: ......